Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline CC panoramic x-ray E-19
Technical Manual
3
CALIBRATING PANORAMIC SENSOR HEAD
Make sure that the unit is in the panoramic mode and that the primary slot number 0 is selected
(the horizontal sector display will remain empty).
The calibration must be performed with 2x2 (high resolution), 3x3 (enhanced resolution) and 4x4
(normal resolution) binning. With high resolution select a kilovolt value of 60 and a milliampere
value of 10. With enhanced resolution select 60 kV and 7 mA, and with normal resolution select
60 kV and 5 mA. Note that these values must be selected to produce a clear exposure.
Remove the inner cover from tube head assembly, if it is in position. Four screws hold the tube
head cover in position, these are located at the rear of the tube head assembly.
Figure
26
Place the calibration tool to the front of the primary slot.
Figure
27
kV
mA
1 0
6 0
Tube head
assembly
Tube head cover
Primary slot
Calibration tool
Содержание Proline CC
Страница 2: ......
Страница 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 158: ...Chapter E DIMAX3 DIGITAL SYSTEM ADJUSTMENT E 70 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Страница 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 206: ......