Chapter
D - ADJUSTMENT & CALIBRATION
Planmeca Proline CC panoramic x-ray D-13
Technical Manual
If the distances from the middle ball to the center of the tenth ball on both the right and left are
less than 118mm then the patient positioning mechanism is too far forward and must be moved
backwards (Fig. 19).
Figure
19
STA
RT
END
PLAN
ME
CA
No.50971
X-LIN
E
Y-LINE
Less than 118mm
Less than 118mm
Right-hand side
Film