4.5.3. Testing the File Structure
Press
«F1»
to start the memory test. When the test is completed, the results will be
displayed on the screen.
If no errors have been detected, the following message will appear (see figure 4.29):
Figure 4.29
There are three types of testing. Type 1 – checking of the files location table. In case an
error is identified, the corresponding file will be deleted. The top information line will
indicate the number of the deleted files. Type 2 – searching for the damaged clusters,
which do not belong to any file. Such clusters will be erased, and their number will be
indicated in the second line. Type 3 – searching for the damaged files with incomplete
length or files that “overlap” other files, i.e. contain clusters common with other files. The
number of damaged files is indicated in the third line. When the damaged files are
detected, it is necessary to manually determine the names of the files and delete them in
the
Tbl
mode (section 4.5.5).
4.5.4. Determining the Memory Resources
Press
«F2»
to activate the program. When the process is completed, the message
containing information about the amount of memory available (in percents of the total
amount) will be displayed (see figure 4.30).
Figure 4.30
4.5.5. Viewing the Structure Table and Files of all Types
Press
«F3»
to activate the program. The memory structure table will appear. See figure
4.31:
Figure 4.31
The first column (
MemType
) displays the names of the memory structures:
a)
Ch
Pln
:
channel plans (IT-081 Meter, IT-087 or IT-088 Analyzer);
b)
DL Chan
:
channel logs (IT-081 Meter, IT-087 or IT-088 Analyzer );
c)
DL
Spec
:
spectrum logs (IT-081 Meter, IT-087 or IT-088 Analyzer);