Front Panel Operation
1.
Press one of the measurement function buttons to select a
measurement function.
2.
Press SHIFT + NULL buttons to enter MATH submenu.
3.
Use
◁
and
▷
buttons to locate “AVERAGE” submenu, and then
press ENTER to actuate it.
4.
To read the average value, min/max value, total counts and
readings counts, you can press
◁
and
▷
buttons to switch
between those values.
※
Note: Press SHIFT+NULL > OFF or switch to other measurement to
disable this feature. The “MATH” anunnciator on the display indicates the
state of a mathematical feature.
Remote Interface Operation
The following commands show you how to use the Average operation
from your PC terminal.
CALCulate:FUNCtion AVERage
CALCulate:STATe {OFF|ON}
CALCulate:STATe?
CALCulate:AVERage:MINimum?
CALCulate:AVERage:MAXimum?
CALCulate:AVERage:AVERage?
CALCulate:AVERage:COUNt?
4.3.3 Null
Definition
When the null function is enabled, the displayed measurement reading
will show something
difference between the measured input signal
reading and the stored null (also called relative) value. The null (relative)
value is stored in a volatile memory and the value will be cleared when
the multimeter is power-off. Moreover, you can specify or alter a null
value.
The null function is available for all types of measurement except
continuity and diode. When you select the null function for a