![Picotest M3510A Скачать руководство пользователя страница 55](http://html1.mh-extra.com/html/picotest/m3510a/m3510a_user-manual_1555486055.webp)
55
55
55
55
4
10
−
⋅
−
=
αδ
B
8
10
−
⋅
−
=
αβ
C
If you are using SPRTD (Standard Platinum RTD), select SPRTD and
specify the seven coefficients under SPRTD submenu.
The ITS (International Temperature Scale) -90 standard provides two
reference equations for Standard Platinum Thermometers covering
temperature range from 18.8033K to 1234.93K. However, one SPRTD
usually cannot cover the entire range, so the temperature range is
broken up into a few subranges. These subranges depend on the
calibration point of temperature scale and are based on the melting or
triple points of various pure substances. For an exact list of the elements
needed and details on RTD calibration, refer to NIST Technical Note 1265
“Guidelines For Realizing the International Temperature Scale of 1990”.
In each subrange, the calibration constants required for that subrange
are listed.
Default
The default sensor type in M3510A/11A is PT100.
How to set up RTD
You can set up the RTD configuration either through the front panel
operation or through the remote interface operation.
Front Panel Operation
Sensor selection
1.
Press TEMP button to select RTD temperature measurement
function.
2.
Press CONFIG button.
3.
Use
◁
and
▷
to locate SENSOR submenu. Press ENTER to select it.
4.
Use
◁
and
▷
to locate your sensor type. Press ENTER to select
your desired sensor type.
Choosing
USER
takes you to a menu where you can specify factors used