![Picotest M3510A Скачать руководство пользователя страница 42](http://html1.mh-extra.com/html/picotest/m3510a/m3510a_user-manual_1555486042.webp)
42
42
42
42
[SENSe:]FUNCtion “<function>”
For example:
FUNCtion[1/2] “VOLTage:DC”
FUNCtion[1/2] “VOLTage:AC”
FUNCtion[1/2]”CURRent:DC”
FUNCtion[1/2] “CURRent:AC”
FUNCtion[1/2] “FREQunecy”
FUNCtion[1/2] “FREQunecy:VOLT”
FUNCtion[1/2] “PERiod”
FUNCtion[1/2] “PERiod:VOLT”
FUNCtion[1/2] “FREQunecy:CURR”
FUNCtion[1/2] “PERiod:CURR”
FUNCtion[1/2] “RESistance”
FUNCtion[1/2] “FRESistance”
FUNCtion[1/2] “CAPacitance”
FUNCtion[1/2] “TEMPerature”
FUNCtion[1/2] “TCOuple”
FUNCtion[1/2] “NONE”
FUNCtion[1/2]?
When enabling the 2
nd
function, you can still set the range and NPLC
individually. The setting procedures are as usual.
To set 2
nd
and NPLC through the remote interface, use the following
commands:
[SENSe:]<function>:RANGe {<range> | MINimum | MAXimum}
[SENSe:]<function>:RANGe:AUTO {OFF | ON}
[SENSe:]<function>:NPLCycles
{0.001|0.006|0.02|0.06|0.2|0.6|1|2|10|100|MIN|MAX}
To read and fetch values from the DMM, use the following commands:
READ[1|2]?
FETCh[1|2]?