How to use 2ND measurement
1.
Press one of the measurement function keys to select a primary
measurement.
2.
Press SHIFT+ENTER to enter 2ND submenu.
3.
Use
◁ or ▷
button to locate a desired secondary measurement, and
press ENTER to select it.
4.
Observe and take readings on the display. The upper row display is
the primary measurement and the lower row display is the secondary
measurement.
※
Note: To simultaneously measure the DCI & DCV from the same input
source, you have to use three leads. The voltage and current
measurements must share the same common lead at the LO terminal.
However, this application will cause an inaccurate measurement due
to the leads combining with a small amount of internal resistance in
the DMM and forming an IR which directly and annoyingly affects the
measurement. For instance, 30 m
Ω
of lead resistance will cause
approximate 90 mV error at 3 A.
※
Note: To execute the 2nd function with DCI, please adjust the range
at 1A or up.
Remote Interface Operation
To set 2
nd
through the remote interface, use the following commands:
These commands are used for selecting two functions on the 2
nd
applications. 1 stands for the selected function which will act on the main
display. 2 stands for the selected function which will act on the secondary
display.