![Picotest M3510A Скачать руководство пользователя страница 105](http://html1.mh-extra.com/html/picotest/m3510a/m3510a_user-manual_1555486105.webp)
105
105
105
105
Appendix
This appendix contains the performance specifications of the
M3510A/11A. It covers the AC, DC, Resistance, Temperature, and
Frequency/Period characteristics under a variety of conditions. It also
contains the general characteristics and accuracy calculations for your
convenience. A lot of efforts are made to make sure these specifications
serve your needs for production, engineering and/or research purposes.
All specification applies to M3510A/11A unless noted otherwise.
A. Specification List
1
DC CHARACTERISTICS
Function
Range
Resolution
1 Year Accuracy
2
DCV
100.0000 mV
0.1 µV
0.0080+0.0045
1.000000 V
1.0 µV
0.0090+0.0010
10.00000 V
10 µV
0.0120+0.0020
100.0000 V
100 µV
0.0120+0.0020
1000.000 V
1 mV
0.0200+0.0030
DCI
10.00000 mA
10 nA
0.050+0.020
100.0000 mA
100 nA
0.050+0.010
1.000000 A
1 µA
0.150+0.020
3.00000 A
3
10 µA
0.200+0.030
10.00000 A
10 µA
0.250+0.050
2WΩ
4
/4WΩ
100.0000Ω
100 µΩ
0.020+0.005
1.000000 KΩ
1 mΩ
0.020+0.002
1
The specifications are under 2-hour warm-up condition with the
setting 10 PLC, and they’re relative to the calibrator specifications in
PICOTEST.
2
± (% of r % of range), (23
℃
℃
℃
℃
± 5
℃
℃
℃
℃
)
3
The M3511A doesn’t support the selection of the range 3A.
4
The Null function must be used when the 2WΩ is selected.