PLC is suggested to select. However, the NPLC setting is used for DCV,
DCI, Ω2 & Ω4 functions only. The lowest PLC on each digit range is the
rapidest selection, such as 0.001 PLC at 4 1/2 digits, 0.2 PLC at 5 1/2
digits and 10 PLC at 6 1/2 digits. The following table 4-3 shows the
relationship between the integration time and the measurement
resolution.
Table 4-3
Integration Time (PLC)
Resolution (with Full-Scale)
0.001
0.0003
0.006
0.0002
0.02
0.0001
0.06
0.00005
0.2
0.00001
0.6
0.000005
1
0.000003
2
0.000002
10
0.000001
100
0.0000003
In addition, the resolution setting applies to all math operations under
the selected measurement function. The selected value is stored in a
volatile memory and the setting is only valid for the present
measurement functions. You can select different resolution for different
measurement functions.
Default
The default value for the resolution is 6 1/2 digits with 10 PCL and the
default value will be restored when the meter has been turned off or after
a remote interface reset. The manual selected value for the resolution
setting is stored in a volatile memory and the setting is only valid for the
present measurement function. Please refer to
Table 2-1
on
page 23.
How to set the resolution
The resolution can be set either through the front panel or the remote
interface operation.