2.
Press CONFIG, and then use
◁
and
▷
to toggle between options.
3.
To configure unit or thermocouple type, press ENTER on “UNIT” and
“TYPE” respectively.
Real Temperature Setup
1.
Selet a type and matched an adaptor correctly.
2.
Press SHIFT+ TEMP to choose thermocouple function.
3.
Press CONFIG, and then use
◁
and
▷
to locate “REAL”, then press
ENTER to check the present temperature.
4.
Press ESC to reach the upper layer.
5.
Press ◁
and
▷
to locate “RJUNCTION”, then do the prcedure ENTER
> REAL to execute the function.
Simulated Temperature Setup
1.
Selet a type and matched an adaptor correctly.
2.
Press SHIFT+ TEMP to choose thermocouple function.
3.
Press CONFIG, and then use
◁
and
▷
to locate “SIMULATED”, then
press ENTER to define a proper value after calculating.
4.
Press ESC to reach the upper layer.
5.
Press ◁
and
▷
to locate “RJUNCTION”, then do the prcedure
ENTER > SIMULATED to execute the function.
Remote Interface Operation
Use the following commands to set up the thermocouple configuration.
SENSe
:UNIT {Cel|Far|K}
SENSe
:UNIT?
SENSe:
TCOuple:TYPE {E|J|K|N|R|S|T}
SENSe:
TCOuple:SIMulated {<
value
>|MINimum|MAXimum}
4.1.8 Remote Interface Selection
The M3510A/11A provides a standard USB and supports an optional
GPIB/RS-232 interface, but only one interface can be activated at a time.
If you are using GPIB, you must set the address for the multimeter. You