C.8 The System-Related Commands
FETCh?
Transfer readings stored in memory by the INITiate command to output buffer
where you are able to read them into your bus controller.
READ?
Change the state of the triggering system from the “idle” state to
“wait-for-trigger” state. The meter will start to make measurements when a
required triggering condition is met after the READ? command is received.
Measurement readings are sent to the output buffer immediately.
DISPlay {OFF|ON}
Turn off or on the display.
DISPlay?
Query the display setting. Returns “0” (OFF) or “1” (ON).
DISPlay:TEXT <
quoted string
>
Show a message on the front panel display. The allowed message can be up to
16 characters in the lower row display; any additional characters will be
truncated.
DISPlay:TEXT?
Query the message sent to the front panel display.
DISPlay:TEXT:CLEar
Clear the message shown on the front panel display.
SYSTem:BEEPer
Issue a single beeping tone.
SYSTem:BEEPer:STATe {OFF|ON}
Disable or enable the beeper.
SYSTem:BEEPer:STATe?
Query the beeper mode. Returns “0” (OFF) or “1” (ON).