![Picotest M3510A Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/picotest/m3510a/m3510a_user-manual_1555486019.webp)
19
19
19
19
Figure 2-5
【
Step 6
】
Insert the voltage setting selector container back into the socket.
2.1.2.2 To change the fuse
Warning! Before replacing the power-line fuse, make sure the
multimeter is disconnected from the AC power. You must be qualified
personnel to perform this action.
Warning! For continued protection against fire or instrument
damage, only replace with fuse of the same type and rating noted on the
rear panel. If the instrument repeatedly blows fuses, locate and correct
the cause of the trouble before replacing the fuse.
Verify that the power-line fuse is good. Replace the fuse if it is damaged.
Follow the steps below to change the fuse.
【
Step 1
】
Verify that the meter is disconnected as shown in
Figure 2-6
.