Front Panel Operation
1.
First select your desired measurement function by pressing one of
the function buttons located on the first row of your meter's front
panel.
2.
Press DIGITS button to select a desired resolution for your
measurement.
※
Note: When following the Way One to set the resolution, your options
with each press on the same button are 4.5, 5.5 and 6.5. And under the
DCV, DCI, Ω2 & Ω4 functions, the NPCL just can be set by pressing
CONFIG > NPLC > ENTER >
◁
and
▷
>
Target PCL.
Remote Interface Operation
Use the following commands on your PC terminal to set the resolution for
your measurement.
CONFigure:<function> <range>,<resolution>
MEASure:<function>? <range>,<resolution>
SENSe:<function>:RESolution <resolution>
You can specify the resolution in the same units as the measurement
function, not in number of digits. For instance, specify the resolution in
vots and the frequency in hertz. Below are some examples.
CONF:VOLT:DC 10,0.001
(4 1/2 digits on the 10 Vdc range)
MEAS:CURR:AC? 1,1E-6
(6 1/2 digits on the 1 A range)
CONF:FREQ 1 KHz,0.1 Hz
(1000 Hz input, 0.1 Hz resolution)
VOLT:AC:RES 0.05
(50 mV resolution on the ac function)
4.1.4 Threshold Resistance (Continuity)
Definition
When testing continuity, the beeper goes off when the measured
resistance is less than the threshold resistance. The threshold resistance
can be set to any value between 1Ω and 1000Ω.