Chapter 4 Treatment

With HIVAMAT

®

200 you can perform treatments with the applicator.

The effects of the massage with HIVAMAT

®

200 develop only if the applicator is continuously moved.

Moist skin of the patient may also interfere with the effectiveness of the massage. Therefore, the area re-
ceiving the massage ought to be dried or possibly powdered.

CAUTION

Before each treatment, adjust the intensity to match the respective conditions and pay atten-
tion to the contraindications (refer to

Contraindications

on page 3)!

4.1 Preparations and Attachment

Basically, treatment is done using an applicator. This mode of treatment is also suitable for self-treatment

How to Prepare the Applicator

(1)

Plug both of the

Spiral Lead Universal (Ref.-No 00356)

items into the

connector sockets <9>

.

Since the instrument operates in biphase mode, you don't have to pay attention to the polarity.

(2)

Connect the applicator and the

Titanium Neutral Element (Ref.-No. 00351)

with the remaining

ends of the spiral leads.

(3)

Let the patient grip the titanium neutral element (or grip it yourself in case of a self-treatment).

(4)

Continue with the steps described under

How to Prepare Treatment

on page 13.

How to Prepare Treatment

(1)

Before treatment, prepare the patient's skin area to be treated by drying or powdering.

(2)

Start up the instrument as described under

How to Start up the Instrument in Mains Mode

on page 9

.

(3)

Turn the

intensity control <6>

to zero.

(4)

Select the frequency as desired using the

frequency selector <3>

. For the effects of the different

frequencies refer to

Massage Frequency

on page 14.

(5)

Select the massage mode using the

mode selector <4>

. For the effects of various mode adjustments

refer to

Instrument Modes

on page 14.

(6)

Set the duration of the massage using the

timer <5>

. For the selection of the massage duration refer

to

Massage Duration

on page 14.

(7)

Start the massage by turning up the intensity using the

intensity control <6>

(also refer to

Massage

Intensity

on page 15).

Important

You have to choose the intensity according to the minimum dose principle, i.e. in the lower
effective range!

The applicator should stay in contact with the patient's skin and should be moved on the

4 Treatment

PHYSIOMED ELEKTROMEDIZIN AG

HIVAMAT

®

200

13

Содержание HIVAMAT 200

Страница 1: ...0899 GB O P E R A T I N G I N S T R U C T I O N S HIVAMAT 200...

Страница 2: ...PHYSIOMED ELEKTROMEDIZIN AG The therapy mode has been patented worldwide HIVAMAT 200 is made in Germany in compliance with the quality requirements of EN ISO 13485 2003 AC 2007 EN ISO 9001 2008 and th...

Страница 3: ...lector 4 5 2 1 5 Timer 5 5 2 1 6 Intensity Control 6 6 2 1 7 Intensity Display 7 6 2 1 8 Turn Down Indicator 8 6 2 1 9 Connector Sockets 9 6 2 1 10 Patient Indicator 10 7 2 1 11 Charge Display 11 7 Ch...

Страница 4: ...Service 16 A 2 Cleaning and Disinfection 16 A 3 Service Life and Disposal 16 A 4 Electromagnetic Compatibility 17 A 5 Training 17 A 6 Technical Data 17 A 6 1 Environment Conditions 18 Appendix B Scop...

Страница 5: ...the technical specifications of IEC 601 VDE 0750 and is assigned to class IIa according to the Council Directive 93 42 EEC concerning Medic al Devices It is not intended for operation in explosion haz...

Страница 6: ...ment functions are automatically checked during a self test routine The intention of treatment with HIVAMAT 200 is not to replace existing therapeutic methods On the contrary it is aimed at supporting...

Страница 7: ...n to be treated Its frequency changes between 5 250 Hz according to the selected indication Due to the movement of the applicator a vibrating or pumping effect with deep impact is induced in the patie...

Страница 8: ...indicators and controls of HIVAMAT 200 The numbers in angle brackets refer to the Instrument Overview on page 2 2 1 1 Mains Module 1 The mains module 1 with mains supply fuses and mains switch is sit...

Страница 9: ...he remaining duration of the massage in minutes The flashing decimal point indicates that the timer is running The plus and minus keys serve for lengthening or shortening the mas sage time in 1 minute...

Страница 10: ...ity control 6 should be turned to zero This is the case following self test routine after time has elapsed and when attempting to extend massage duration with the intensity turned high 2 1 9 Connector...

Страница 11: ...rge display 11 consists of two diodes The upper green diode is continuously lit when the battery is being charged It starts flashing after feed end The lower yellow diode flashes if the battery is und...

Страница 12: ...for a longer period of time Important In order to ensure a long battery life the batteries must be charged completely when first charged The first charging procedure should not be interrupted How to C...

Страница 13: ...mains operation while the battery is charged NOTICE If the unit is not used for a long period of time please charge the battery once a month This will help to avoid complete discharge 3 2 Preparation...

Страница 14: ...trument The instrument will automatically carry out a self test routine during which all functions and output values are checked All displays light up an automatic self test routine is performed by th...

Страница 15: ...e up the oscillator head in the respective size The oscillator head is screwed into the handle To remove the head from the handle you have to rotate it to the left Important Any damage to the oscillat...

Страница 16: ...automatic self test routine or during operation a special error code is displayed on the display In this case massage is automatically interrupted To check whether the instrument operates properly ex...

Страница 17: ...the titanium neutral element or grip it yourself in case of a self treatment 4 Continue with the steps described under How to Prepare Treatment on page 13 How to Prepare Treatment 1 Before treatment...

Страница 18: ...y can be described as follows High frequency approximately 80 200 Hz High frequencies are most effective for local areas near the surface Medium frequency approximately 25 80 Hz Medium and low frequen...

Страница 19: ...in wide movements Compression and velocity of the massage are adapted to intensity and frequency Start massage from the proximal edge of the area to receive the massage in direction of flow Only when...

Страница 20: ...mmendations see Manufacturer s Recommendations on page 21 A 2 Cleaning and Disinfection Clean the device and the accessories regularly with an aldehyde based disinfectant Switch off the instru ment by...

Страница 21: ...nstrument is not absolutely essential If there is a need however continued training can be provided Please contact the PHYSIOMED ELEKTROMEDIZIN AG A 6 Technical Data Protection Class acc to VDE 0750 I...

Страница 22: ...n Ambient temperature 10 C 40 C Relative humidity 30 75 Air pressure 800 1 060 hPa Storage and Transport Ambient temperature 10 C 40 C Relative humidity 10 90 Air pressure 500 1 060 hPa PHYSIOMED ELEK...

Страница 23: ...ment 1 00577 Transportation bag Basic 1 00899 Operating Instructions English 1 01522 Therapy Manual English 1 B 2 Additional Accessories The following additional accessories are available for HIVAMAT...

Страница 24: ...Ref No Designation 01001 Trolley Universal 01005 Trolley Universal III PHYSIOMED ELEKTROMEDIZIN AG Scope of Delivery and Accessories 20 HIVAMAT 200...

Страница 25: ...entary Documents C 1 Manufacturer s Recommendations 0 1 2 3 4 3 3 3 051 061 071 3 3 3 081 3 3 091 95686 5 95686 5 2 09 1 0 3 6 1 6 8 09 51 09 61 3 8 5 Supplementary Documents PHYSIOMED ELEKTROMEDIZIN...

Страница 26: ...on 4 inflammations acute 3 instrument modes 14 instrument description 2 instrument errors 12 instrument overview 2 intensity control 6 intensity display 6 M mains module 4 maintenance 16 manufacturer...

Страница 27: ...W weight 17 Index PHYSIOMED ELEKTROMEDIZIN AG HIVAMAT 200 23...

Отзывы: