Now, there are two possible results:
•
If the
patient indicator <10>
ceases to light, the spiral lead is in order.
•
If the
patient indicator <10>
is lit or flashes, the spiral lead is defective and should be replaced.
3.5 Hand Applicator
The hand applicator consists of the following parts:
Hand Applicator (Exploded View)
(1) Handle (Ref.-No. 00379)
(2) Cover
(3) Oscillator Plate
(4) Membrane
(5) Clamp Ring
NOTICE
(2) - (5) make up the oscillator head in
the respective size.
The oscillator head is screwed into the handle. To remove the head from the handle, you have to rotate it
to the left.
Important
Any damage to the oscillator membrane might lead to harmless (but unpleasant) sensations
on the skin.
Replace the membrane immediately if damaged!
How to Replace the Membrane
(1)
Take off the clamping ring and remove the membrane.
(2)
Insert the new membrane with the foil facing down loosely into the clamping ring, so that the mem-
brane foil protrudes laterally.