   TROUBLESHOOTING

 

 

If the unit malfunctions, check out the following before calling for service. 
 

Malfunction Check/Remedy 

The unit does not operate 
at all 

 The unit is not connected to the power supply properly. 

 The power switch is not turned on. 

(MPR-S163-PE / MPR-S313-PE only) 

 The capacity and voltage of power supply is not sufficient. 

 There is a power failure. 

 The circuit breaker on the supply circuit is activated. 

 A breaker of 15 A or more is recommended. 

The alarm is activated at 
start-up 

 The alarm status is kept until the chamber temperature reaches the 

set temperature. 

 The alarm buzzer recovers after 30 minutes even if the alarm buzzer 

is silenced by pressing the alarm buzzer stop key (BUZZER). 

The alarm is activated 
during operation

 

 The door was kept opened for a long time. 

 The door is opened. 

 An error code is displayed on the temperature display. (page 23) 

 Contact our sales representative or agent. 

Noisy 

 The floor is not sturdy. 

 The installation site is not level. 

 The unit is tilted. 

 The cabinet contacts the surrounding wall. 

When the refrigerator   
does not get cold 
enough

 

 The door is frequently opened. 

 The unit is in direct sunlight. 

 The ventilation around the unit is blocked. 

 There is a nearby heat source. 

 The ambient temperature is too high. 

 The allowable ambient temperature is between -5 

o

C and 35 

o

C. 

 The storage items are too much. 

 The air exhaust vent is blocked up with storage items. 

 The access port is not covered. 

 The access port should be covered with the insulation and rubber 

caps when no use. 

 
 

   DISPOSAL OF UNIT

 

 

 

WARNING 

If the unit is to be stored unused in an unsupervised area for an extended period 

ensure that children 

do not have access and doors cannot be closed completely.

 

The disposal of the unit should be accomplished by appropriate personnel.

   

Always remove 

doors

 to prevent accidents such as suffocation. 

 

  Decontamination of chamber

 

 
Before disposing the pharmaceutical refrigerator with biohazardous danger, decontaminate the 
pharmaceutical refrigerator to the extent possible by the user. 

Содержание MPR-S163 Series

Страница 1: ...es Please read the operating instructions carefully before using this product and keep the operating instructions for future use See page 32 for all model numbers Operating Instructions Pharmaceutical...

Страница 2: ......

Страница 3: ...TTING P 17 Setting of high temperature alarm P 17 Setting of low temperature alarm P 18 SETTING OF DELAY OF DOOR ALARM P 19 SETTING OF ALARM BUZZER RECOVERY P 20 OPERATION CHECK AFTER RECOVERY P 21 RE...

Страница 4: ...ent of performance or functions Contact our sales representative or agent if any page of the operating instructions is lost or page order is incorrect Contact our sales representative or agent if any...

Страница 5: ...her person Precautions are illustrated in the following way WARNING Failure to observe WARNING signs could result in a hazard to personnel possibly resulting in serious injury or death CAUTION Failure...

Страница 6: ...wer supply outlet is not grounded it will be necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may...

Страница 7: ...e electric shock or fire Do not position this unit and the other unit so that it is difficult to operate the disconnection of the power supply plug Failure to disconnect the power supply plug may caus...

Страница 8: ...operation after power failure or turning off of power switch The stored items may be damaged due to the change of setting Be careful not to tip over the unit during movement to prevent damage or injur...

Страница 9: ...C Maximum relative humidity 80 for temperature up to 31 o C decreasing linearly to 50 relative humidity at 40 o C Mains supply voltage fluctuations up to 10 of the nominal voltage Transient overvoltag...

Страница 10: ...8 Rear side REFRIGERATOR COMPONENTS 2 17 19 18 13 1 MPR S163 10 8 3 2 6 4 9 11 12 5 7 18 19 17 MPR S313 1 3 2 4 6 5 7 10 11 12 13 8 9 16 MPR S163 PE MPR S313 PE only 15 14 6 Power supply cord...

Страница 11: ...r a temperature recorder Location for an optional temperature recorder page 30 10 Key lock Insert the key and turn to the clockwise 180o with pressing into the head of lock when lock the doors 11 Shel...

Страница 12: ...Digit shift key Pressing this key at setting mode makes the digit to be inputted on the temperature display to change Pressing this key for 5 seconds at temperature display mode causes the setting mod...

Страница 13: ...n or noise WARNING Install the unit on a sturdy floor If the floor is not strong enough or the installation site is not adequate this may result in injury from the unit falling or tipping over Select...

Страница 14: ...floor Ensure the unit is horizontally Fig 1 3 Fixing the unit Two fixtures are attached to the rear of the frame Fix the frame to the wall with these fixtures and rope or chain Fig 2 Note Contact our...

Страница 15: ...page 14 The factory setting of chamber temperature is 5 o C Important When the power supply has been turn off wait 5 minutes or more after recovery of power supply connect the power supply plug to th...

Страница 16: ...e is displayed Turn on the power switch MPR S163 PE MPR S313 PE only 2 Press the set key SET The current setting is displayed and the first digit blinks 3 Set the temperature to 4 with the digit shift...

Страница 17: ...t key SET is not memorized DEFROSTING The following 2 types of defrosting methods is provided for the refrigerator Both of them are controlled automatically Cycle defrosting The chamber temperature is...

Страница 18: ...ift key and numerical value shift key When pressed the settable digit is shifted When pressed the figure of settable digit changes 4 Press the set key SET The current setting is displayed and the firs...

Страница 19: ...numerical value shift key for about 5 seconds F00 is displayed and the first digit blinks 3 Press the numerical value shift key and scroll the figure to 1 When pressed the figure of settable digit ch...

Страница 20: ...is displayed and the first digit blinks 3 Press the numerical value shift key and scroll the figure to 2 When pressed the figure of settable digit changes 4 Press the set key SET The current setting i...

Страница 21: ...rding to the condition of use Description of operation Key operated Indication after operation 1 The current chamber temperature is displayed 2 Press the numerical value shift key for about 5 seconds...

Страница 22: ...20 minutes from 30 minutes Following shows an example Change the setting as necessary Description Operation Key operated Indication after operation 1 The current chamber temperature is displayed 2 Pre...

Страница 23: ...ERMINAL OPTION A remote alarm terminal is mounted on the battery mounting box MPR 48B1 Optional component The alarm is outputted from this terminal Contact capacity is DC 30 V 2 A Contact output betwe...

Страница 24: ...1 installed Power failure alarm Power failure The power supply cord is disconnected The power switch is off MPR S163 PE MPR S313 PE only Power failure alarm MPR 48B1 installed Alarm indicator blinks I...

Страница 25: ...02 and 50 is displayed alternately The defrost sensor is disconnected Alarm indicator blinks E03 and chamber temp are displayed alternately The defrost sensor is short circuited Alarm indicator blinks...

Страница 26: ...iluted detergent always wipe it off with a wet cloth Then wipe off the cabinet or accessories with a dry cloth Important Never pour water onto or into the unit This may cause electric shock or failure...

Страница 27: ...sy The floor is not sturdy The installation site is not level The unit is tilted The cabinet contacts the surrounding wall When the refrigerator does not get cold enough The door is frequently opened...

Страница 28: ...method of disposal German Benutzerinformationen zur Entsorgung von elektrischen und elektronischen Ger ten private Haushalte Dieses Symbol auf Produkten und oder begleitenden Dokumenten bedeutet dass...

Страница 29: ...ter les autorit s locales ou votre revendeur afin de conna tre la proc dure d limination suivre Spanish Informaci n sobre la eliminaci n para los usuarios de equipos el ctricos y electr nicos usados p...

Страница 30: ...pe a informa es sobre o m todo de elimina o correcto Italian Informazioni per gli utenti sullo smaltimento di apparecchiature elettriche ed elettroniche obsolete per i nuclei familiari privati Questo...

Страница 31: ...act op met de lokale overheid of uw leverancier en vraag wat de juiste verwijderingsmethode is Swedish Information om kassering f r anv ndare av elektrisk elektronisk utrustning privata konsumenter Om...

Страница 32: ...our sales representative or agent for the attachment of a temperature recorder Temperature recorder Recorder fixing MTR 0621LH MPR S30 MTR G04A or MTR G04C MPR S7 INTERFACE BOARD OPTION Two kinds of...

Страница 33: ...Hermetic type Output 160 W Evaporator Fin and tube type Condenser Wire and tube Refrigerant R 407D Defrosting New cycle defrost and forced defrost Defrost heater 87 W 91 W USA only 101 W Other area Dr...

Страница 34: ...duct name Pharmaceutical refrigerator MPR S313 Model number MPR S313 PA MPR S313 PK MPR S313 PT MPR S313 PR MPR S313 PE Temperature control range Between 2 o C and 14 o C ambient temp 5 o C to 35 o C...

Страница 35: ...he unit Unit interior No contamination Decontaminated Contaminated Others 3 Instructions for safe repair maintenance disposal of the unit a The unit is safe to work on b There is some danger see below...

Страница 36: ...Printed in Japan LDCL035400 0 S0418 0 1 1 1 Sakada Oizumi machi Ora gun Gunma 370 0596 Japan PHC Corporation 2018...

Отзывы: