25
TROUBLESHOOTING
If the unit malfunctions, check out the following before calling for service.
Malfunction Check/Remedy
The unit does not operate
at all
□
The unit is not connected to the power supply properly.
□
The power switch is not turned on.
(MPR-S163-PE / MPR-S313-PE only)
□
The capacity and voltage of power supply is not sufficient.
□
There is a power failure.
□
The circuit breaker on the supply circuit is activated.
→
A breaker of 15 A or more is recommended.
The alarm is activated at
start-up
□
The alarm status is kept until the chamber temperature reaches the
set temperature.
□
The alarm buzzer recovers after 30 minutes even if the alarm buzzer
is silenced by pressing the alarm buzzer stop key (BUZZER).
The alarm is activated
during operation
□
The door was kept opened for a long time.
□
The door is opened.
□
An error code is displayed on the temperature display. (page 23)
→
Contact our sales representative or agent.
Noisy
□
The floor is not sturdy.
□
The installation site is not level.
□
The unit is tilted.
□
The cabinet contacts the surrounding wall.
When the refrigerator
does not get cold
enough
□
The door is frequently opened.
□
The unit is in direct sunlight.
□
The ventilation around the unit is blocked.
□
There is a nearby heat source.
□
The ambient temperature is too high.
→
The allowable ambient temperature is between -5
o
C and 35
o
C.
□
The storage items are too much.
□
The air exhaust vent is blocked up with storage items.
□
The access port is not covered.
→
The access port should be covered with the insulation and rubber
caps when no use.
DISPOSAL OF UNIT
WARNING
If the unit is to be stored unused in an unsupervised area for an extended period
ensure that children
do not have access and doors cannot be closed completely.
The disposal of the unit should be accomplished by appropriate personnel.
Always remove
doors
to prevent accidents such as suffocation.
Decontamination of chamber
Before disposing the pharmaceutical refrigerator with biohazardous danger, decontaminate the
pharmaceutical refrigerator to the extent possible by the user.