24
ROUTINE MAINTENANCE
CAUTION
Always disconnect the power supply to the unit prior to any repair or maintenance
of the unit in
order to prevent electric shock or injury.
Ensure you do not inhale or consume medication or aerosols
from around the unit at the time of
maintenance. These may be harmful to your health.
Cleaning of cabinet
■
Clean the unit once a month. Regular cleaning keeps the unit looking new.
■
Use a dry cloth to wipe off small amounts of dirt on the outside and inside of the unit and all accessories.
If the outside panels are dirty, clean them with a diluted neutral dishwashing detergent. Wipe off the
condensation on the glass or exterior of the cabinet with a dry soft cloth. (Undiluted detergent can
damage the plastic components. For the dilution, refer to the instruction of the detergent.)
After the cleaning with the diluted detergent, always wipe it off with a wet cloth. Then wipe off the
cabinet or accessories with a dry cloth.
<Important>
■
Never pour water onto or into the unit. This may cause electric shock or failure.
■
Do not use a brush, an acid, a thinner, laundry soap, a powder detergent, boiling water
for
cleaning.
These cause damage of painted surface or failure of plastic and rubber components. Also, do
not wipe the plastic and rubber components by a volatile material.
Cleaning of evaporating tray
The evaporating tray is placed in the bottom rear of
the unit. Clean the evaporating tray twice or three
times a year.
1.
Remove the fixing plate by pulling the clip toward
front. (Fig.1)
2.
Pull out the evaporating tray. (Fig.2)
3.
Dispose remaining water in the evaporating tray.
4.
Clean the evaporating tray with dilute neutral
dishwashing detergent and then rinse away the
detergent with water. Do not use hot water.
5.
After the cleaning, replace the fixing plate to its
original position, and fix it by the clips.
Evaporating tray
Fig.2
Clip
Fig.1
Fixing plate
Содержание MPR-S163 Series