20
SETTING OF ALARM BUZZER RECOVERY
The alarm buzzer recovers and notifies the abnormality again after preset period (alarm buzzer recovery
time) if the same alarm status is continued after the alarm buzzer is stopped by pressing the alarm buzzer
stop key (BUZZER). Always set the alarm buzzer recovery time to avoid the false recognition of alarm
status.
Settable range of alarm buzzer recovery time:
10 minutes interval between 10 and 60 minutes (setting; between 010 and 060) or
No recovery (setting; 000)
Initial setting (at factory): 30 minutes
The alarm buzzer never recovers once the alarm buzzer is silenced by pressing the alarm buzzer stop
key (BUZZER) if the alarm buzzer recovery time is set to 000 (no recovery). Nevertheless, the alarm
buzzer sounds when other alarm status is detected.
Example: Changing the alarm buzzer recovery time to 20 minutes from 30 minutes.
Following shows an example. Change the setting as necessary.
Description Operation
Key operated
Indication after operation
1 -----
-----
The current chamber temperature is
displayed.
2
Press the numerical value shift
key for about 5 seconds.
F00 is displayed and the first digit
blinks.
3
Set the figure to F25 with the digit
shift key and the numerical value
shift key.
When pressed, the settable digit
is shifted.
When pressed, the figure of settable
digit changes.
4 Press the set key.
SET
The current recovery time is
displayed. The second digit blinks.
5
Scroll the figure to 020 with the
numerical value shift key.
When pressed, the figure of settable
digit changes.
6 Press the set key.
SET
The setting is memorized and the
current chamber temperature is
displayed.
■
The setting mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation. In this case, any setting before pressing set key (SET) is not memorized.
■
Do not set the alarm buzzer recovery time during the alarm status.