21
OPERATION CHECK AFTER RECOVERY
After recovery from a power failure, the unit will resume operation automatically with the setting (chamber
temperature, lock of temperature setting, alarm temperature, alarm buzzer recovery time) before power
failure. Accordingly, there is no need for re-set however, always check the running status after recovery.
■
The setting is memorized by nonvolatile memory during power failure.
<Attention>
■
After recovery from a power failure, always check the unit starts the operation normally with the setting
before power failure.
REMOTE ALARM TERMINAL (OPTION)
A remote alarm terminal is mounted on the battery mounting box (MPR-48B1). (Optional component)
The alarm is outputted from this terminal. Contact capacity is DC 30 V, 2 A.
Contact output:
between COM. and N.O.
between COM. and N.C.
At
normal
Open
Close
At
abnormal
Close
Open
Note:
■
The buzzer is silenced by pressing alarm buzzer stop key (BUZZER) on the control panel during alarm
condition. (A remote alarm is continuing the operation.)
The buzzer will be activated again after certain suspension if the alarm condition is continued.
■
The alarm is actuated when the power supply plug is disconnected from the outlet or the power switch
is OFF for MPR-S163-PE and MPR-S313-PE only.
Battery switch
Remote alarm terminal
Space for interface board
Battery mounting box