10
REFRIGERATOR COMPONENTS
Control panel components
1. Door check indicator (DOOR):
The red LED lamp is lit when the door is opened.
(After 2 minutes the door check indicator lights; the alarm buzzer is activated to notice the door opening.)
2. Alarm indicator (ALARM):
The red LED lamp blinks during an alarm status. (page 22, 23)
3. Temperature display:
Normally, the current chamber temperature is displayed in the temperature
display. When the high/low temperature alarm or 0
o
C alarm is activated, the current chamber
temperature blinks (refer to page 22). When the self-diagnostic function detects an abnormality, the
error cord is displayed (refer to page 23).
4. Numerical value shift key ( ):
Pressing this key for 5 seconds at temperature display mode causes
the setting mode. Pressing this key at setting mode makes the figure on the temperature display to
change.
5. Digit shift key ( ):
Pressing this key at setting mode makes the digit to be inputted on the
temperature display to change. Pressing this key for 5 seconds at temperature display mode causes the
setting mode for key lock.
6. Set key (SET):
Pressing this key at temperature display mode enables the temperature setting.
( page 14) Pressing this is at the end of setting mode memorizes the setting.
7. Alarm buzzer stop key (BUZZER):
Pressing this key silences the alarm buzzer when the alarm
buzzer sounds. But the remote alarm is not canceled.
8. LED light key (LIGHT):
Turn on the LED light by pressing this key. To turn off the LED light, press
this key again.
If it is turned on by LED light key, it will be turned off automatically after setting time.
LED light turns off when the door is opened and then closed.
■
“temperature display mode”: the status which the temperature display shows the current chamber
temperature
■
“setting mode”; the status which the temperature display is ready to be entered after pressing
numerical value shift key for about 5 seconds.
8
7
6
5
2
1
3
4