16
LED LIGHTING TIME SETTING
The LED light is automatically turned off after setting time when the LED light is turned on by pressing the
LED Light key. The method of change of lighting time is as follows.
Initial setting (at factory): 10 minutes
Example: Changing the LED lighting time to 5 minutes from 10 minutes
Following shows an example. Change the setting as necessary.
Description Operation
Key operated
Indication after operation
1 -----
-----
The current chamber temperature is
displayed.
2
Press the numerical value shift
key for about 5 seconds.
F00 is displayed and the first digit
blinks.
3
Set the figure to F45 with the digit
shift key and numerical value shift
key.
When pressed, the settable digit
is shifted.
When pressed, the figure of settable
digit changes.
4 Press the set key.
SET
The current setting is displayed and
the first digit blinks.
5
Set the figure to 005 with the digit
shift key and the numerical value
shift key.
When pressed, the settable digit
is shifted.
When pressed, the figure of settable
digit changes.
6 Press the set key.
SET
The setting is memorized and the
current chamber temperature is
displayed.
■
The setting mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation. In this case, any setting before pressing set key (SET) is not memorized.
■
The LED lighting time can be changed from 0 minute to 15 minutes.
■
The LED light keeps off when the lighting time set to 000 in procedure 5.