17
ALARM TEMPERATURE SETTING
Setting of high temperature alarm
The abnormal temperature rise is notified by the blink of alarm indicator and display of chamber
temperature, and the alarm buzzer (15 minutes after blink) when the high temperature alarm is set.
Always set the high temperature alarm to avoid the damage of storage items due to the temperature rise.
Settable range of high temperature alarm:
Between the set temperature of chamber plus 2
o
C and
the set temperature of chamber plus 14
o
C
Initial setting (at factory): the set temperature of chamber plus 5
o
C
Example: Changing the high temperature alarm to 3
o
C from 5
o
C plus the set temperature of chamber.
Following shows an example. Change the setting as necessary.
Description Operation
Key operated
Indication after operation
1 -----
-----
The current chamber temperature
is displayed.
2
Press the numerical value shift
key for about 5 seconds.
F00 is displayed and the first digit
blinks.
3
Press the numerical value shift
key and scroll the figure to 1.
When pressed, the figure of settable
digit changes.
4 Press the set key.
SET
The current setting is displayed and
the first digit blinks.
5
Scroll the figure to 003 by using
the digit shift key and the
numerical value shift key.
When pressed, the settable digit
is shifted.
When pressed, the figure of settable
digit changes.
6 Press the set key.
SET
Alarm temperature is memorized
and the current chamber
temperature is displayed.
■
The setting mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation. In this case, any setting before pressing set key (SET) is not memorized.
<Important>
■
The alarm may be activated after defrosting or when too much items are stored in the chamber,
depending on the setting of high temperature alarm. This is not a malfunction. The alarm is cancelled
automatically when the chamber temperature reaches the set temperature.