3.8.6 Wiring for CP operating mode, ion counter, with HV 702
HV 702
HV
+
HV
-
CD
RF A
RF B
FA
IS
EP
QC 700
CP
QMA
CP 400
HV+
QC
HV-
9
10
SEM 217
8
Fig. 41:
Wiring for CP operating mode, ion counter, with HV 702
This wiring applies to applications with HV 702 and operating modes 2 and 3.
Position
Cable
Length (m)
Order number
8
HV connection HV 702 – CP 400
3
10
BG541978 -T
BG541979 -T
9
10
Connection QC 700 – CP 400
3
10
BG448134 -T
BG448199 -T
Tbl. 13:
Wiring for CP operating mode, ion counter, with HV 702, operating modes 2 and 3
3.9 Identifying the product
You will need all the data from the rating plate to safely identify the product when communicating with
Pfeiffer Vacuum.
► To ensure clear identification of the product when communicating with Pfeiffer Vacuum, always
keep all of the information on the rating plate to hand.
3.10 Scope of delivery
● QMG 700 HiQad (individual components, dependent on configuration)
● Operating instructions
Unpacking and checking product
1. Unpack the product.
2. Remove the transport fasteners, transport protection etc.
3. Store the transport fasteners, transport protection etc. in a safe place.
4. Ensure that no parts are damaged.
Product description
40/70