![Pfeiffer Vacuum HIQUAD QMG 700 Скачать руководство пользователя страница 38](http://html1.mh-extra.com/html/pfeiffer-vacuum/hiquad-qmg-700/hiquad-qmg-700_operating-instructions-manual_1534983038.webp)
3.8.4 Wiring for 90°-SEM operating mode with HV 702
HV 702
HV
+
HV
-
CD
RF A
RF B
FA
IS
EP
QMH 4xx
EP 2
SEM
QMA
EP
HV+
HV -
7
8
SEM 217
12
11
11
Fig. 39:
Wiring for 90°-SEM operating mode with HV 702
This wiring applies to applications with HV 702 and operating mode 0.
Position
Cable / jump plug
Length (m)
Order number
7
Connection EP 422 (2) – QMH 4xx
0.85
- (mounted on EP 422)
8
HV connection HV 702 – QMA 4x0 (HV–)
3
10
BG541978 -T
BG541979 -T
11
HV jump plug (SHV type)
-
B4728891B9
12
EP jump plug (TNC type)
-
B4728138BC
Tbl. 11:
Wiring for 90°-SEM operating mode with HV 702, operating mode 0
Product description
38/70
Содержание HIQUAD QMG 700
Страница 70: ...BG5400 ed B Date 2210 P N BG5400BEN...