![Pfeiffer Vacuum HIQUAD QMG 700 Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/pfeiffer-vacuum/hiquad-qmg-700/hiquad-qmg-700_operating-instructions-manual_1534983035.webp)
3.8.1 Basic wiring for Faraday operating mode
RF A
RF B
FA
IS
EP
QC 700
IS 716
QMH 4xx
QMA
EP 1
FARAD
EP 2
SEM
QMH
RF+
FA
RF -
1)
QMA
6
5
4
1
2
3
1)
Fig. 36:
Basic wiring for Faraday operating mode
This wiring applies to applications with HV 701 and HV 702.
Position
Cable
Length (m)
Order number
1
Connection IS 716 – QMA 4x0
3
10
BG548082 -T
BG548083 -T
2
Connection QC 700 – QMH 4xx
Extension cable
3
7
- (mounted on QMH 4xx)
BG448175 -T
3
Connection EP 422 (1) – QMH 4xx
0.85
- (mounted on EP 422)
4
1)
HF connection QMH 4xx – QMA 4x0
0.7
BG448295 -T
5
1)
6
FA connection QMH 4xx – QMA 4x0
0.7
BG541962 -T
1)
Cross cables 4 and 5 if required according to the test report.
Tbl. 8:
Basic wiring for Faraday operating mode