Accupik Accessory . 81
Alignment Procedure
NOTE: You must have a crimped sample pan in a crucible in the autosampler in order to align the
Accupik properly. Use the Volatile Sample Pan Sealer Accessory (P/N N5370435) to crimp a
sample pan (P/N N5370550).
1.
When the Accupik is installed it should have at least 5.0mm space between the
crucible handle and needle assembly.
2.
From the Pyris Software TGA 8000 Control Panel, click on the Autosampler Control