16 . TGA 8000 Installation and Hardware Guide 

 

Hazardous Chemical Warnings 

Before using any chemicals or solvents with the instrument, you should be thoroughly familiar 
with all hazards and safe handling practices. Observe the manufacturer's recommendations 
for use, storage, and disposal. These recommendations are normally provided in the material 
safety data sheets (MSDS) supplied with the solvents. 

 

 

We use the term 

CAUTION

 to inform you about situations that could result in 

serious damage to the instrument or other equipment. 

Nous utilisons le terme 

AVERTISSEMENT

 pour vous informer de situations 

pouvant entraîner des blessures corporelles si les précautions ne sont pas 
respectées. 

 

  

 

WARNING: Some chemicals used with this instrument may be hazardous or 
may become hazardous after completion of an analysis.

 The responsible body 

(e.g., Lab Manager) must take the necessary precautions to ensure that the 
surrounding workplace and instrument operators are not exposed to hazardous 
levels of toxic substances as defined in the applicable Material Safety Data Sheets 
or OSHA, ACGIH, or COSHH documents. Venting for fumes and disposal of waste 
must be in accordance with all national, state, and local health and safety 
regulations and laws. 

AVERTISSEMENT: Certains produits chimiques utilisés avec cet instrument 
peuvent être dangereux ou risquent de devenir dangereux une fois l'analyse 
terminée

. L'organisme responsable (p. Ex., Le responsable du laboratoire) doit 

prendre les précautions nécessaires pour s'assurer que le lieu de travail 
environnant et les exploitants d'instruments ne sont pas exposés à des niveaux 
dangereux de substances toxiques tels que définis dans les fiches de données de 
sécurité ou les documents OSHA, ACGIH ou COSHH. L'évacuation des fumées et 
l'élimination des déchets doivent être conformes à toutes les réglementations et lois 
nationales, nationales et locales en matière de santé et de sécurité. 

Some definitions of terms used above are:

 

 

Responsible body:

 "Individual or group responsible for the use and maintenance of 

equipment, and for ensuring that operators are adequately trained" [per IEC 1010-1, 
Amendment 2]. 

 

Operator:

 "Person operating equipment for its intended purpose" [per IEC 1010-1, 

Amendment 2]. 

 

OSHA:

 Occupational Safety and Health Administration (United States) 

 

ACGIH:

 American Conference of Governmental Industrial Hygienists 

 

COSHH:

 Control of Substances Hazardous to Health (United Kingdom) 

Installation Category 

This instrument is able to withstand transient overvoltage according to Installation Category II 
as defined in IEC 1010-1. 

Содержание TGA 8000

Страница 1: ...TGA 8000 Installation and Hardware Guide THERMAL ANALYSIS...

Страница 2: ...fitness for a particular purpose PerkinElmer shall not be liable for errors contained herein for incidental consequential damages in connection with furnishing performance or use of this material Copy...

Страница 3: ...20 Order Spares and Accessories 24 Installing the TGA 8000 25 TGA 8000 Installation Summary 26 Unpack the TGA 8000 27 Set Up TGA 8000 System Components 29 Install the Purge Gases 29 Connect the TGA 80...

Страница 4: ...75 Installation 75 Connecting to the Chiller 76 Setting up the Chiller refer to Circulator User Manual for more details 78 Setting the Coolant Temperature 79 Accupik Accessory 80 Mounting the Accupik...

Страница 5: ...5 Safety and Regulatory Information...

Страница 6: ...The following safety conventions are used throughout We use the term WARNING to inform you about situations that could result in personal injury if precautions are not followed Nous utilisons le terme...

Страница 7: ...e danger of burns if precautions are not followed due to the presence of heat in this area of the instrument There may also be hot surfaces that can be easily touched AVERTISSEMENT Surface chaude This...

Страница 8: ...isfaisant et s r de l instrument il est essentiel que le conducteur de terre de protection le conducteur vert jaune du cordon d alimentation secteur soit connect une vraie masse lectrique Toute interr...

Страница 9: ...Service on electrical components should be performed only by a qualified PerkinElmer service engineer Be sure the power cord is the correct one for your laboratory The TGA 8000 power cord must be rate...

Страница 10: ...message appears It is now safe to shut off your computer Never press the Reset button on the computer if the Pyris software appears to malfunction Press the Ctrl Alt Del keys simultaneously and select...

Страница 11: ...0 while it is on Wait until the power is off and the furnace cooling fan has stopped spinning before moving the instrument Do not remove cables from the analyzer while it is on Do not touch the openin...

Страница 12: ...f equipment Never operate equipment with the grounding connector disconnected Disconnect supply cord before operating AVERTISSEMENT La continuit du circuit de mise la terre est essentielle au bon fonc...

Страница 13: ...bels 13 Warning Label behind the Furnace CAUTION HOT SURFACE which indicates that the furnace at the center of the ring may be hot ATTENTION SURFACE CHAUD qui indique que le four au centre de l anneau...

Страница 14: ...er operate equipment with the grounding connector disconnected Disconnect supply cord before operating AVERTISSEMENT La continuit du circuit de mise la terre est essentielle au bon fonctionnement de l...

Страница 15: ...or relocated or separated by a greater distance from the interfering source Consult PerkinElmer for additional assistance WARNING If the heating of materials could lead to the liberation of hazardous...

Страница 16: ...al state and local health and safety regulations and laws AVERTISSEMENT Certains produits chimiques utilis s avec cet instrument peuvent tre dangereux ou risquent de devenir dangereux une fois l analy...

Страница 17: ...n conductive POLLUTION occurs Occasionally however a temporary conductivity caused by condensation must be expected South Korea Ingress Protection The TGA 8000 is rated IP20 as defined in IEC 60529 St...

Страница 18: ...recycling and recovery programs vary by regulatory authority at your location Contact your local responsible body for example your laboratory manager or authorized representative for information rega...

Страница 19: ...19 Prepare the Laboratory...

Страница 20: ...measures the deflection with an optical sensor and uses current to return the beam to its original position The amount of current necessary to maintain the system in the null state is directly propor...

Страница 21: ...ng a constant heating rate or a more complex temperature program Below are some techniques that can be used to obtain the most accurate data in a minimum amount of time with a TGA 8000 Sample Preparat...

Страница 22: ...they do not react with the furnace materials thermocouple or other analyzer components in which gases come into contact Analyses are done at normal pressure or at reduced pressures In addition sample...

Страница 23: ...n the outlet side for connection to the analyzer s purge gas line CAUTION Maximum pressure should not exceed that listed in the chart below ATTENTION La pression maximale ne doit pas d passer celle in...

Страница 24: ...kinElmer s web based catalog service offers a full selection of high quality thermal analysis supplies To place an order request a free catalog or ask for information If you are located within the U S...

Страница 25: ...25 Installing the TGA 8000...

Страница 26: ...y Unpack the TGA 8000 Set Up the TGA 8000 System Components Connect the TGA 8000 System Components Configure the TGA 8000 Install the Furnace Hangdown Wire and Balance Tare Weight Calibrate the TGA 80...

Страница 27: ...ed to automate the running of samples If you ordered a TGA with an autosampler it is shipped with the autosampler already installed When you unpack the box check all the parts included against the shi...

Страница 28: ...res pour le faire en toute s curit Le banc doit tre proximit NE soulevez PAS l appareil par le d me sup rieur Ce d me prot ge le m canisme de l quilibre et n est pas viss 6 One person should put one h...

Страница 29: ...balance chamber The recommended flow rate is 20 100 ml min for the balance purge and 10 50 ml min for the sample gas ATTENTION Le d bit de purge du syst me doit toujours tre sup rieur au d bit de gaz...

Страница 30: ...rn past fingertight NOTE If you are splitting the System Purge to supply also Sample Gas A cut a short length of 1 8 in Teflon tube and connect as described above by using the T splitter Connect the l...

Страница 31: ...l sera n cessaire d utiliser un syst me d extraction des fum es Connecting Reactive Gas The TGA 8000 has a reactive gas assembly with which a reactive gas can enter the sample area The reactive gas en...

Страница 32: ...lowing order Computer TGA 8000 Printer 8 Turn on the purge gas supply and the water for the TGA 8000 cooling system if applicable or any other accessories for the cooling device you are using Configur...

Страница 33: ...lyzer For the TGA 8000 the following accessories are available Cooling Jacket Accessory Accupik GMD 8000 Firmware Version Click the Firmware Version button to display the version of firmware in the an...

Страница 34: ...osampler Adjustment Configure the iOS App for the TGA 8000 This procedure will help guide you in configuring the TGA 8000 for use with an iPad iPhone The interface is designed so the Pyris Light app c...

Страница 35: ...an connect to When complete click OK 3 Select Remote Control from the Pyris Tools menu 4 In the Remote Control dialog the Local IP address and Port of the PC that the app will connect with is displaye...

Страница 36: ...Click the New Analyzer button 7 Enter the IP Address Port Number and Name of the Analyzer enter any name then click Next This displays the Settings screen which displays the General settings Display g...

Страница 37: ...37 8 Click Accept these settings can be changed at any time This displays the Main Screen displaying a list of all Analyzers 9 Select your analyzer and the connection will be made This is confirmed w...

Страница 38: ...the Remote Control dialog in Pyris which displays the Local IP address and Port of the iPad or iPhone Connection permissions can be controlled with all devices within this dialog Start Stop button St...

Страница 39: ...8 after it is released from the balance plate Failure to do this can result in damaging the balance mechanism 3 Observe the level on the balance plate near the flange of the anti convection tube c 4 C...

Страница 40: ...e platinum hangdown ribbons are extremely fragile Be careful not to exert force on the loops of the hangdown ribbon or on the hangdown wire after it is attached Handle tweezers carefully when placing...

Страница 41: ...ound The wire should be centered in the iris 2 If it is not centered then the balance mechanism must be adjusted Adjusting the Balance Mechanism If the hangdown wire is not centered in the iris then y...

Страница 42: ...operation of the autosampler ATTENTION Si le TGA 8000 est d plac ou install pour la premi re fois la configuration initiale des fils d quilibrage et de d connexion doit tre effectu e avant l talonnage...

Страница 43: ...e tray Ensure the hook of the hangdown wire is Aligned 2 3mm below the apex of sample pan handle right image below The hangdown wire is centered within the furnace tube right image below Aligned on th...

Страница 44: ...pt For this calibration an empty sample pan must be present on the hangdown wire before starting During the calibration the sample pan should hang on the centerline of the resting position on the tray...

Страница 45: ...Initial Setup for the Balance and Hangdown Wires 45 6 Make sure an empty sample pan is present on position 1 Select carousel location as 1 and click Load Sample button...

Страница 46: ...n needs only to be performed once provided the full calibration was not set to defaults The Temperature and Baseline Calibration may be re done to optimize the TGA for optimum temperature and weight l...

Страница 47: ...isplay should look as below with the value in the Measured mg box reading 100 000 If it indicates another value you must restore the default weight calibration using the Calibration Restore Weight Cal...

Страница 48: ...w calibration value to the analyzer and save the calibration file 11 Click the Close button to exit the Calibration window The weight calibration is complete Performing the Temperature Calibration The...

Страница 49: ...ceramic pan and then cover them with thin coating of alumina powder 4 With a non autosampler system attach the pan onto the hangdown wire raise the furnace and observe the angle of the pan bale stirr...

Страница 50: ...ly place the magnet P N N5370466 around the furnace tube then raise the furnace 10 Rotate the magnet until the pan rotates from the magnetic field to its pre magnet angle Make sure the sample pan is n...

Страница 51: ...point will be used to determine one of the points for each weight loss step 6 Select Weight Loss curve so the derivative curve goes to the background 7 Select Onset from the Calc menu 8 Move the two...

Страница 52: ...the analyzer and save the current calibration file SN_weightcal_st tg8c 16 Click Close to exit the Calibration window The temperature calibration is now complete For the case where you are calibrating...

Страница 53: ...utosampler Control dialog box to load a sample 5 Click Begin Calibration An estimate of the amount of time needed to perform the calibration is displayed and it begins to count down 6 When the calibra...

Страница 54: ...minimum and maximum temperatures between which the baseline will be determined e g set minimum to 50 C and maximum to 1150 C 6 Select a scanning Rate C min that you will be using for scanning analyse...

Страница 55: ...55 TGA 8000 Hardware Maintenance...

Страница 56: ...d other accessories such as hangdown wires and thermocouples may need to be replaced NOTE The TGA 8000 case is painted metal The exterior surfaces may be cleaned with a soft cloth dampened with a mild...

Страница 57: ...nace Tube In order to clean the furnace tube you must first remove the tube from the furnace Clean the furnace tube using one of the following methods 1 Soak and scrub the furnace tube in a strong det...

Страница 58: ...emoving the furnace components To perform this procedure the furnace must be in the lowered position AVERTISSEMENT Avant d effectuer cette proc dure arr tez le syst me et retirez l alimentation secteu...

Страница 59: ...he Tygon exhaust tube from the furnace tube Lay the furnace tube on the bench top 5 Remove the inner Pyrex furnace lining 6 Remove the platinum furnace shield by sliding it off the furnace Do not defo...

Страница 60: ...ere is a locating pin on the mounting assembly that fits into the hole on the bottom of the furnace to assist alignment 3 Place the furnace assembly onto the mounting assembly 4 Place the knurled furn...

Страница 61: ...12 Place the clamp around the base of the furnace tube and top of the furnace tube base The red handle should be placed so that it is facing the cooling housing The exhaust tube should be to the righ...

Страница 62: ...Check the position of the stirrup with respect to the tip of the thermocouple 4 Adjust the height of the furnace using the furnace adjust height knob located below the furnace adjustment ring Turn th...

Страница 63: ...now exposed 3 Open the iris all the way by pushing the arm clockwise 4 Loosen the screws on the two stands that hold the iris in place 5 Move the iris around until the hangdown wire is centered in th...

Страница 64: ...NTION Ne pas tirer le tube du four tr s rapidement ou angle pour viter de briser la tige mince sur le four Replacing the Furnace Tube This procedure is performed after removing the furnace tube 1 Care...

Страница 65: ...austic acid solution Wear protective gloves and safety glasses and perform this procedure in a fume hood AVERTISSEMENT H2SO4 concentr et K2CrO7 est une solution d acide caustique tr s forte Porter des...

Страница 66: ...e thermocouple avec pr caution car cela pourrait l endommager Si le thermocouple est vieux us ou cass remplacez le P N N5321002 2 Remove the tip of the thermocouple from inside the furnace by pulling...

Страница 67: ...and raise or lower the thermocouple such that it is just touching the bottom of the cap Remove the cap 5 Plug the plastic thermocouple connector plug on the base of the thermocouple into the base of...

Страница 68: ...68 TGA 8000 Installation and Hardware Guide...

Страница 69: ...xert force on the loop of the balance ribbon or on the hangdown wire after it is attached Handle tweezers with extra care when placed in the vicinity of the platinum balance ribbon the hangdown wire o...

Страница 70: ...d to change the weight carefully remove the present one using tweezers Pick up the new tare weight with tweezers and carefully hang it from the loop of the hangdown ribbon on the right side of the bal...

Страница 71: ...all so that the ribbon can be removed For an autosampler system the orientation of the platinum wire loop is important If affects the position of the hook which is critical in sample pan loading and u...

Страница 72: ...om the bottom end of the capillary tube 7 Once the ball is free from the capillary tube clean off any residual rubber cement Use as many drops of acetone as needed Acetone that remains trapped in the...

Страница 73: ...not been disturbed If so repeat steps 14 and 15 17 Locate and open the tube of rubber cement P N 09923480 Apply a tiny amount of cement to the end of the wire rubber cement applicator P N N5376742 To...

Страница 74: ...re Device for an autosampler system N5320016 Aluminum crucible liner N5376732 Kit of 400 aluminum liners N5370492 Stainless steel crucible liner N5376749 Kit of 400 stainless steel crucible liners N53...

Страница 75: ...e 1 roll 09920073 Quick connect stem 2 09940294 Nylon tie wrap 6 N5370333 Glass tube 1 Installation To install the cooling jacket accessory follow the steps below 1 Switch your TGA on If it is already...

Страница 76: ...er of the chiller s pump Distilled water is recommended for operation at temperatures between 15 C and 90 C For operation at 20 C to 100 C 50 ethylene glycol in water is recommended ATTENTION Un fluid...

Страница 77: ...other hose coupling into the INLET connection on the chiller and tighten with a suitable wrench 8 Push a second piece of hose onto the INLET coupling and secure with a hose clamp 9 Connect the hose f...

Страница 78: ...ing the Power switch on the front of the control unit 4 When switching on the chiller for the first time select the display language by rotating the Select Set control on the front panel and then pres...

Страница 79: ...to maintain a very constant flow rate of coolant The liquid should be free of air bubbles Changes in the liquid flow rate or the presence of air bubbles will reduce the quality of the measurement sig...

Страница 80: ...then enters the furnace The puncturing of the sample pan provides an additional level of protection against volatilization of samples Now you can program volatile sample analysis for unattended overn...

Страница 81: ...to align the Accupik properly Use the Volatile Sample Pan Sealer Accessory P N N5370435 to crimp a sample pan P N N5370550 1 When the Accupik is installed it should have at least 5 0mm space between...

Страница 82: ...ng Depth buttons are grayed out inactive 3 Click Change to set the Piercing Mode to ON The Piercing Position and Piercing Depth buttons become active 4 Make sure that a crimped sample pan use the Accu...

Страница 83: ...6 Loosen the black thumbscrew adjustment knob to carefully move the stem so that the needle assembly contains the needle tip is slightly above the sample in the crucible The space between the needle...

Страница 84: ...in order to maintain the 5 mm space between the needle s tip and the top of the pan 1 Remove the Accupik to replace the needle 2 Using a screwdriver loosen the screw that is attached to the metal shaf...

Страница 85: ...85 TGA 8000 Autosampler...

Страница 86: ...nneur automatique sans l aide d un technicien de maintenance WARNING Do not remove the gripper or covers No user serviceable parts are inside the autosampler Refer all servicing to a qualified service...

Страница 87: ...or the Lowered position The autosampler is in the Safe position 2 Start the play list Your play list may have an explicit Load Sample command or if using a Sample Group the entry Sample List implies L...

Страница 88: ...ound the sample and the sheath gas if used baths the furnace liner that is the gas goes between the Pyrex furnace tube and the liner not in the furnace Sample Loading Before loading the samples into t...

Страница 89: ...e Weigh System dialog box appears The system automatically starts the program to weigh each sample in the list After the last sample if there was a missing sample encountered a message is displayed Th...

Страница 90: ...he system purge and sample gas tubing are properly connected Turn on the gases adjust the pressures The recommended flow rate for the system purge is 20 100 ml min For the sample gas it should be betw...

Страница 91: ...ing a TGA 8000 Autosampler Playlist 91 6 While taring the TGA Tare Weigh System screen is displayed 7 The software will tare the crucibles it finds and populate the Sample List in the Sample Group for...

Страница 92: ...ed is displayed You can select the method after the samples have been weighed Click Weigh List The TGA Tare Weigh System screen appears and the system begins to weigh the samples listed in the Sample...

Страница 93: ...le to add to the list 11 Save the play list by selecting Save Player from the File menu 12 Start playback of the play list by clicking on the Start at Top button or the Start at Current Item button on...

Страница 94: ...The black area of the autosampler is anodized aluminum To clean it you can use a cloth dampened with isopropyl alcohol 2 The sample tray is clear anodized aluminum which can be cleaned with isopropyl...