TGA 8000 Furnace Maintenance . 59
2.
Slide the instrument so the furnace is positioned off the edge of the bench
3.
Loosen and remove the Clamp holding the Pyrex furnace tube to the base of the
furnace assembly.
4.
Remove the Pyrex furnace tube. You do not have to disconnect the Tygon exhaust
tube from the furnace tube. Lay the furnace tube on the bench top.
5.
Remove the inner Pyrex furnace lining.
6.
Remove the platinum furnace shield by sliding it off the furnace. Do not deform the
shield.
7.
Remove the plate (slotted ring) sitting on the spring.
8.
Remove the spring
9.
Unscrew the large knurled furnace nut to completely remove the furnace from the
mount
10.
Remove the clip that is underneath the nut.
11.
Lift the furnace assembly off of the mounting.