Environmental Requirements for the TGA 8000 . 15
Environmental Requirements for the TGA 8000
The TGA 8000 has been designed for indoor use only and should not be operated in an explosive
environment. The TGA 8000 operates most efficiently under the following conditions:
•
Ambient temperature range of 10°C – 35°C (50°F – 95°F). The instrument will remain
safe when operated between 5°C and 40°C (41°F to 104°F).
•
Ambient relative humidity between 20% and 80% without condensation.
•
Altitude no more than 2000 m.
•
Clean area free of dust, smoke, vibration, and corrosive fumes.
•
On a bench area out of direct sunlight.
•
Away from heating or cooling units or ducts.
•
The instrument must be positioned so the power switch on the rear panel can be
easily reached by the operator.
•
There must be an adequate and stable power source for all system components.
•
The TGA 8000 is a highly sensitive, precision laboratory instrument. The environment
in which it is used must be free of radio frequency interference sources which may
affect the performance of the instrument. If performance is affected by RF fields, the
instrument should be reoriented or relocated, or separated by a greater distance from
the interfering source. Consult PerkinElmer for additional assistance.
WARNING:
If the heating of materials could lead to the liberation of hazardous
gases, the use of a fume extraction system will be required.
AVERTISSEMENT:
Si le chauffage des matériaux peut entraîner la libération de
gaz dangereux, il sera nécessaire d'utiliser un système d'extraction des fumées.
WARNING:
There may be possible hazards of explosion, implosion, or the
release of toxic or flammable gases arising from the materials being heated.
AVERTISSEMENT:
Il peut y avoir des risques d'explosion, d'implosion ou de
dégagement de gaz toxiques ou inflammables provenant des matières chauffées.
WARNING:
Do not mount the instrument of a surface of flammable material.
AVERTISSEMENT:
Ne pas monter l'instrument d'une surface de matériau
inflammable.
CAUTION:
Before using any cleaning or decontamination method, except that
recommended in this guide, the user should check with PerkinElmer Instruments
that the proposed method will not damage the instrument.
ATTENTION:
Avant d'utiliser toute méthode de nettoyage ou de décontamination,
à l'exception de celle recommandée dans ce guide, l'utilisateur doit vérifier auprès
de PerkinElmer Instruments que la méthode proposée n'endommagera pas
l'instrument.
Содержание TGA 8000
Страница 1: ...TGA 8000 Installation and Hardware Guide THERMAL ANALYSIS...
Страница 5: ...5 Safety and Regulatory Information...
Страница 19: ...19 Prepare the Laboratory...
Страница 25: ...25 Installing the TGA 8000...
Страница 55: ...55 TGA 8000 Hardware Maintenance...
Страница 68: ...68 TGA 8000 Installation and Hardware Guide...
Страница 85: ...85 TGA 8000 Autosampler...