TGA 8000 Furnace Maintenance . 61
9.
Place the slotted plate on top of the spring. Make sure the plate is straight and the
slot aligns with the thermocouple.
10.
Place the quartz insert over the furnace and seat it on the plate.
11.
Place the Pyrex furnace tube over the furnace so that it rests on the O-ring on the
top of the furnace tube base. Position the exhaust tube so that it is left of center
when the furnace is out of the cooling position.
12.
Place the clamp around the base of the furnace tube and top of the furnace tube
base. The red handle should be placed so that it is facing the cooling housing. The
exhaust tube should be to the right of the clamp. Note the position of the exhaust
tube and clamp when the furnace is in the Cooling position in the right-hand photo
below.
13.
Plug in the furnace power cord to its connector on the left side of the TGA. There are
male and female guides on the connector that allow the connector to go on in only
one way.
After installing the standard furnace, you will have to adjust it with respect to the ball joint,
hangdown wire, and sample pan.