92 . TGA 8000 Installation and Hardware Guide
8.
Remove the sample tray from the autosampler and place the samples in the
crucibles. Carefully place the ring back on to the autosampler. You can also remove
each crucible from the tray individually and load the sample into the crucible and
return it to the tray.
Now you can have the system weigh each sample in the list. You can also choose to
have each sample weighed right before the run starts. To weigh all the samples
before starting the play list, highlight the Sample List line. A message informing you
that a method has not been selected is displayed. You can select the method after
the samples have been weighed. Click Weigh List. The TGA Tare/Weigh System
screen appears and the system begins to weigh the samples listed in the Sample
List. When complete, the weights are displayed, along with the tare weights, in each
sample line.
9.
Next, you need to add some items to the Data Analysis List. If you have not entered
a method for the Sample List, however, you cannot fill in the Data Analysis List. To
enter a method, highlight the Sample List line. Type in the name of the method in
the Method Name field or click the Browse button and find and select the method
you want to use. Edit the method's program and initial state parameters by clicking
the Edit Method button. When finished editing, close the Method Editor by clicking
Содержание TGA 8000
Страница 1: ...TGA 8000 Installation and Hardware Guide THERMAL ANALYSIS...
Страница 5: ...5 Safety and Regulatory Information...
Страница 19: ...19 Prepare the Laboratory...
Страница 25: ...25 Installing the TGA 8000...
Страница 55: ...55 TGA 8000 Hardware Maintenance...
Страница 68: ...68 TGA 8000 Installation and Hardware Guide...
Страница 85: ...85 TGA 8000 Autosampler...