Initial Setup for the Balance and Hangdown Wires . 43
2.
In the Pyris Software TGA 8000 screen (located at the top of the screen) click
Calibrate Autosampler from the Calibrate menu.
The Autosampler Calibration dialog box is displayed:
3.
Calibrate the tray height by clicking on the Tray Vertical button and follow the on-
screen prompt. Make sure an empty sample pan is present in position 1 in the
sample tray. Ensure the hook of the hangdown wire is:
•
Aligned 2-3mm below the apex of sample pan handle (right image below)
•
The hangdown wire is centered within the furnace tube (right image below))
•
Aligned on the centerline of the apex. (left image below).
NOTE: If the hangdown is not on the centerline of the apex, the unit will have to be adjusted using
the leveling feet. Either raise or lower the foot nearest the furnace to ensure the hangdown
wire is on center with the apex. If this procedure is done, re-check to ensure the iris is not
contacting the hangdown wire (see Adjusting the Convection Iris section).