Set Up TGA 8000 System Components . 31
Exhaust Gas Tubing
3.
Connect the barbed fitting (P/N 09920026) to the exhaust port on the rear panel.
4.
Insert the end of a piece of 0.25-inch Tygon tubing and to the barbed fitting
connector on the back of the instrument labeled EXHAUST. The tube should be
placed beneath a hood vent if you expect to have noxious exhaust fumes. Otherwise,
it can just hang from the instrument.
WARNING:
If the heating of materials could lead to the liberation of hazardous
gases, the use of a fume extraction system will be required.
AVERTISSEMENT:
Si le chauffage des matériaux peut entraîner la libération de
gaz dangereux, il sera nécessaire d'utiliser un système d'extraction des fumées.
Connecting Reactive Gas
The TGA 8000 has a reactive gas assembly with which a reactive gas can enter the sample
area. The reactive gas enter through the rear of the TGA
When using corrosive reactive gas, connect it to a connector on the rear of the balance
dome. The assembly is underneath the balance dome and goes down into the upper furnace
tube (ball joint).
Connect the TGA 8000 Components
The computer and optional printer should already be connected together and the Pyris
software already installed. To complete the connection of the system components, the TGA
8000 has to be connected to the computer.
CAUTION:
Make sure all components are turned off and unplugged before you begin.
ATTENTION:
Assurez-vous que tous les composants sont éteints et débranchés
avant de commencer.
Содержание TGA 8000
Страница 1: ...TGA 8000 Installation and Hardware Guide THERMAL ANALYSIS...
Страница 5: ...5 Safety and Regulatory Information...
Страница 19: ...19 Prepare the Laboratory...
Страница 25: ...25 Installing the TGA 8000...
Страница 55: ...55 TGA 8000 Hardware Maintenance...
Страница 68: ...68 TGA 8000 Installation and Hardware Guide...
Страница 85: ...85 TGA 8000 Autosampler...