Prima 320 Advance
86
User Manual
Appendix
3. Safety guide
WARNING
1. This module provides data for exhaled CO2 and respiration rate
only. The data should only play a supporting role in a diagnosis.
Clinical signs and symptoms must be checked before final
diagnosis.
2. Do not reuse disposable sampling pipes.
CAUTION
1. This module is to be used by trained professionals. Read the
manual before using the module.
2. A single-use water trap must not be reused, to maintain
accuracy and prevent damage to the module. The unit must not
be shared between patients to prevent cross-infection.
3. Dispose of components in accordance with hospital regulations
and in an environmentally safe manner.
4. To prevent damage and maintain accuracy always empty the
water trap when it is nearly full. Check the level once a week at
least.
5. The sampling tubes must not be bent. This will prevent pump
overload and maintain accuracy of measurement.
6. Repairs must be made only by Penlon-trained engineers, or by
the manufacturer.
7. The manufacturer will not be held responsible if the operator
uses the unit incorrectly. The module must only be used as
stated in the manufacturer’s instructions.
8. Do not use the module for the measurement of exhaled
gases. Exhaled moisture may cause measurement errors, and
moisture accumulation may reduce module life.
9. Changes in temperature may affect measured data. Always use
in stable temperature environments.
10. Any obstruction to the flow of the sample gas (e.g. tubing
severely bent, contaminants blocking the sampling tube, filters
blocked) may cause inaccurate measurements and damage. An
obstruction lasting more than 20 seconds will trigger a pump
shut down.
11. System risk detection conforms to EN1441: 1997 and EN62366
2008.
12. A pipeline leak will seriously affect the accuracy of
measurement data and waveform shape.
13. Excess humidity can affect measurement accuracy.
4. Connection
Follow the instructions supplied by the manufacturer.
Installation – water trap
Note that the CO2 module is installed inside the machine.
The water trap mounting is on the side of the machine.
Attach the water trap to the mounting and then connect the CO2
sampling components as illustrated.
Connect the sample tube to the water trap sample port.