![Penlon Prima 320 Advance Скачать руководство пользователя страница 85](http://html1.mh-extra.com/html/penlon/prima-320-advance/prima-320-advance_user-manual_1523898085.webp)
Prima 320 Advance
79
User Manual
Maintenance
7.8
Oxygen Concentration Calibration
CAUTION
1. Calibrate the O2 sensor (1) at the same ambient pressure
in which it will be used.
2. Follow your hospital biohazard disposal procedure for the
discarded O2 sensor. Do not incinerate.
1. Remove the O2 sensor.
Check that there is no water build-up on the O2 sensor or
within the port (2) on the side of the absorber.
2. Reinstall the O2 sensor and check that the cable (3) is
connected correctly.
7.8.1 O2 calibration at 21%
CAUTION
1. Calibrate when the measured value of O2 concentration
is outside specification range, or when a new sensor is
fitted.
2. If calibration fails:
a) Check for a technical alarm and then check the
troubleshoot section. Repeat the calibration.
b) In case of repeated calibration failures, fit a new
sensor and repeat the calibration. If it still fails,
contact a Penlon-trained service engineer.
Calibration procedure at 21% O2
1. Select standby mode.
2. Press the hotkeys -> [Calibration] -> [Oxygen Cell Cal]
3. Select [21%] and press confirm to start calibration.
4. Follow the on-screen instructions.
5. After a successful calibration, the screen shows
[Calibration success!].
6. If the message [Calibration Failure!] is displayed, repeat
the calibration.
7.8.2 100% O2 calibration
CAUTION
1. If the calibration fails, check for a technical alarm.
Repeat the calibration.
2. In case of repeated calibration failures, replace the O2
sensor and repeat the 21% O2 calibration.
3. Calibrate at 100% O2 after a 21% O2 calibration is
completed. If it still fails, contact a Penlon trained
engineer.
Calibration at 100% O2:
1. Check that the 21% O2 calibration is completed
successfully and that the O2 Supply Failure alarm is not
triggered.
2. Select standby mode.
1