Prima 320 Advance
76
User Manual
Maintenance
7.4.1 Using disinfection equipment
CAUTION
1. To minimize the impact on the environment, use only
water when disinfecting. The maximum temperature
during disinfection is within the range 90°C to 95°C.
2. Clean or rinse with clean water.
3. Use disinfected equipment for disinfection.
If detergent is used with cleaning and sterilizing, use a
detergent with a pH of <10.
4. Dry the components, assemble and/or store.
7.4.2 Using detergent
CAUTION
1. Use a detergent with a pH of <10.
2. Limit the soak time for the synthetic rubber cover (1) for
the pressure relief valve in the base of the bellows to 15
minutes.
3. Clean or rinse with clean water.
4. Use a mild detergent with a pH of less than 10, for 2%
glutaraldehyde solution).
5. Use clean water to rinse off the residual disinfectant
6. Dry the components, assemble and/or store.
7.4.3 Autoclave
CAUTION
1. Steam autoclave will shorten the service life of the
component.
2. Use a certified process, typically at temperatures of:
a) 121°C (250°F) for 15 minutes, or
b) 134°C (275°F) for 4 minutes.
3. Components can also be autoclaved at a temperature of
134°C (275°F) for 18 minutes.
4. Procedure
a) Clean or rinse with clean water.
b) Dry the components.
c) Use steam autoclave.
d) Dry the components, assemble and/or store.
7.4.4 Display screen
CAUTION
The display screen must not be cleaned with liquid.
1. Wipe with a cotton cloth soaked in soft detergent solution.
2. Wipe off the residual detergent solution with a dry, soft,
lint-free cloth.