Installation and Pre-Use Checks
5.4
Installing the Vaporizer
NOTE
1. Read the vaporizer instruction manual before clinical use.
2. Up to two Selectatec-compatible vaporizers may be fitted.
Install the vaporizer
1. Carefully offer the vaporizer up to the manifold (1).
2. Check that the gas connection ports (2) on the vaporizer
are aligned with the valves (3) on the manifold (1).
3. Carefully lower the vaporizer onto the manifold
4. Turn the locking lever (4) clockwise through and lock the
vaporizer into position by clockwise rotation of the locking
lever (3) through 90°.
NOTE
Do not use excessive force to lock the vaporizer onto the
manifold. Damage to the locking fastener will result.
CAUTION
To prevent damage to the locking shaft (5), check that the gas
connection ports are aligned with the valves on the manifold
(1), and are correctly engaged, before tightening the locking
lever.
5.5
Gas cylinder installation
CAUTION
Open the cylinder valves slowly to avoid damage to the
pressure reducing valve and pressure gauges. Ensure that
valves are at least one full turn open when in use.
1. Fit the gas cylinders to their respective yokes (1):
2. Follow appropriate manual handling guidelines when
lifting.
3. Check that the machine yoke and cylinder faces are dust
free and clean.
The sealing washer (2) must be in position between the
cylinder valve and the yoke.
4. Tighten the yoke securely before opening the cylinder
valve. Dust and dirt presents a fire hazard in the presence
of high pressure gas. Leakage of high pressure gas can
cause serious injury.
5. Open the cylinder valves one at a time and check the
pressure on each gauge.
WARNING
Do not leave a gas cylinder valve open if the pipeline supply
is in use. Cylinder supplies could be depleted, leaving an
insufficient reserve supply in case of pipeline failure.
Gas cylinders must conform with the requirements of
IEC 60601-1.
4
2
2
5
2
2
1
3
1