3.12.5 Synchronised intermittent mandatory
ventilation - Pressure control (SIMV-P)
The ventilator delivers pressure controlled breaths which
are synchronised with the patients inspiratory efforts. The
ventilator also allows spontaneous breaths with pressure
support.
3.12.6 Synchronised intermittent mandatory
ventilation - Pressure-regulated volume
controlled (SIMV-PRVC)
The ventilator delivers synchronised pressure controlled
breaths at the lowest possible pressure in order to deliver
the set tidal volume. The ventilator also allows spontaneous
pressure support breaths.
3.12.7 Spontaneous / Pressure support ventilation
(SPONT/PSV)
The ventilator allows spontaneous breaths and delivers
pressure support at a preset trigger level. In the event of an
apnea condition, the ventilator enters backup mode - either
VCV or PCV, whichever the user has preselected.
3.13 Patient monitoring -
Parameters
Measured parameters are grouped on-screen into two
displays - Values 1 and Values 2.
Values 1
•
Inspiratory tidal volume (VTI)
•
Expiratory tidal volume (VTE)
•
Minute ventilation (MV)
•
Spontaneous minute ventilation (MVspn)
•
Respiratory rate (ftotal)
•
Spontaneous breathing frequency (fspn)
•
Peak airway pressure (Ppeak)
•
Mean airway pressure (Pmean)
•
Inspiratory plateau pressure (Pplat)
•
Positive end expiratory pressure (PEEP)
•
Minimum airway pressure (Pmin)
•
Respiratory ratio (I:E)
•
Airway resistance (Rst)
•
Compliance (Cdyn)
•
Fraction of inspired oxygen - FiO2
•
Peripheral oxygen saturation -SpO2 (optional)
•
Pulse rate (bpm)
•
Perfusion index (PI)
Values 2
•
Fraction of inspired carbon dioxide - FiCO2 (optional)
•
End tidal carbon dioxide - FiCO2 - (optional)
Description