Installation and Pre-Use Checks
Air pipeline test
22. Use the O2 flowmeter control to set a flow of 0.4 L/min.
23. Check that the Air pipeline pressure gauges shows 300 to
600 kPa
24. Disconnect the Air pipeline supply.
5.9.6 Gas cylinder supplies
CAUTION
Open the cylinder valves slowly to avoid damage to the
pressure reducing valve and pressure gauges. Ensure that
valves are at least one full turn open when in use.
1. Fit the gas cylinders to their respective yokes, open the
cylinder valves one at a time and check the pressure on
each gauge.
2. Make sure that the flowmeter controls are turned fully
clockwise until gas supplies are required.
5.9.6.1 Cylinder test - cylinder pressure
1. Disconnect all gas pipeline supplies.
2. Open each cylinder valve and check each cylinder pressure.
3. Replace the cylinder if necessary, refer to section 5.5.
4. Close the cylinder valve.
O2 cylinder high pressure leak test
5. Set the anaesthesia system switch to the OFF position
6. Open the O2 cylinder valve.
7. Make a record of the current cylinder pressure.
8. Close the O2 cylinder valve.
9. Make a record of the cylinder pressure after two minute.
10. If the cylinder pressure decreases more than 700 kPa
(100 psi), install a new cylinder seal.
11. Repeat steps 5 through 9. If the leak continues, contact a
penlon-trained engineer. Do not use the O2 cylinder supply.
12. Reconnect the Oxygen pipeline.
N2O cylinder High pressure leak test
13. Open the N2O cylinder valve, and repeat the leak test
sequence given above (O2 cylinder high pressure leak test,
operations 7 to 12).
14. Reconnect the N2O gas pipeline.
O2 verification
15. Set the anaesthesia system switch to the ON position.
16. Connect the Y-piece to the test lung
17. Press the Standby button.
18. Make sure that the system is in operational mode.
19. Use the O2 Flow control to set an O2 flow of 5 L/min.