Prima 320 Advance
10
User Manual
Description
Backup Time
A fully charged battery will power the anaesthesia machine for
up to 90 minutes, depending on configuration and operation.
For example, frequent use of monitoring modules will shorten
the backup time available.
Battery Recharge
The battery is charged automatically when the machine is
connected to an AC power source.
Recharge time: Approximately 4 hours from total discharge
CAUTION
Check every month and fully charge the battery if necessary.
Battery Life
Battery life depends on frequency and period of use.
We recommend that the battery is replaced every 2 years.
3.7.4 On-screen power status and battery indicators
On-screen power status indicator
AC power is being used.
Battery icon flashing: Battery is charging
Battery icon static: Battery is fully charged
AC power is off, battery power is being used
On-screen battery status indicator
Battery is fully charged
Battery power is being used
Battery power is being used - low capacity
A high level alarm [BATTERY LOW] will be
triggered. Ventilation will still be possible
when a low battery alarm is triggered.
Battery approaching total discharge
When the icon shows empty display sectors,
and the battery total discharge alarm is
triggered, automatic shutdown will occur
unless mains power is restored. See section
6.11.6.
Connect the AC power immediately or use an
alternative ventilation method.
3.7.5 Installing and replacing the battery
WARNING
Battery installation must be undertaken by a trained
engineer, see section 7.11.