Maintenance
7.2
Cleaning and Disinfection
WARNING
1. Disconnect the system from the mains power supply
before maintenance, repairs, cleaning, disinfection and
sterilization.
2. Read the safety information for each cleaning agent, and
the user manual for the disinfection equipment.
3. Wear gloves and safety glasses.
Take care when removing the O2 sensor (1). A damaged
sensor can leak and cause burns (the sensor contains
potassium hydroxide).
4. Reuse of a non-disinfected breathing system or reusable
accessories may cause cross-contamination.
Follow infection control and safety procedures to prevent
cross-infection from components contaminated with
blood and body fluids.
5. To prevent breathing system leaks, avoid damaging the
components when disassembling and reassembling.
Ensure correct installation, especially seals.
Always follow the correct cleaning and disinfection
methods and reassembly procedures, as described in
this manual.
6. Pre-use tests (see sections 5.9 to 5.14) must be
performed before patient use if the anaesthetic machine
has been disassembled for cleaning and disinfection, or
servicing.
CAUTION
1. Clean and disinfect the equipment as required before it is
put into use for the first time.
2. Do not use organic, halogenated, or petroleum based
solvents, anaesthetic agents, glass cleaners, acetone, or
other harsh cleaning agents.
3. Do not use abrasive cleaning agents (such as steel wool,
silver polish or cleaner).
4. Keep all liquids away from electronic parts.
5. Do not allow liquid to enter the equipment housings.
6. Limit the soak time for the synthetic rubber cover for
the pressure relief valve in the base of the bellows to 15
minutes. This applies to all synthetic rubber components.
Swelling or faster aging can occur.
This warning applies to the silicon rubber case for the
pressure relief valve in the base of the bellows (see
section 7.3.1).
7. Cleaning solutions must have a pH of 7.0 to 10.5.