FIELDBUS COMMUNICATION PROTOCOLS
The following table lists the input area registers (prepared by the device and read from the master), common
to all PROFINET, ETHERCAT, ETHERNET/IP, PROFIBUS and DEVICENET fieldbuses.
The registers have a size of 16 bits. The input area is updated to a maximum frequency of 125 Hz (80 Hz
for PROFIBUS fieldbus). The size of the Input area set in the fieldbus master must coincide with the dimension
set in the instrument.
Add
INPUT AREA REGISTER
Note
0
Register Status
See the relevant table
1
Gross weight (MSB)
Value INT. - Most significant word
2
Gross weight (LSB)
Value INT. - Less significant word
3
Net weight (MSB)
Value INT. - Most significant word
4
Net weight (LSB)
Value INT. - Less significant word
5
Peak (MSB)
Value INT. - Most significant word
6
Peak (LSB)
Value INT. - Less significant word
7
Digital inputs
See the relevant table.
8
Digital outputs
9
Monitor register
This value corresponds to the equivalent register in the output area..
10
Net measure weight (MSB)
Value INT. - Most significant word
11
Net measure weight (LSB)
Value INT. - Less significant word
12
Measure code (MSB)
Value INT. - Most significant word
13
Measure code (LSB)
Value INT. - Less significant word
14
Set-Point 1 (MSB)
Value INT. - Most significant word
15
Set-Point 1 (LSB)
Value INT. - Less significant word
16
Set-Point 2 (MSB)
Value INT. - Most significant word
17
Set-Point 2 (LSB)
Value INT. - Less significant word
18
Load cell capacity (MSB)
Value INT. - Most significant word
19
Load cell capacity (LSB)
Value INT. - Less significant word
20
Cell 1 sensitivity
Value INT.
21
Cell 2 sensitivity
Value INT.
22
Cell 3 sensitivity
Value INT.
23
Cell 4 sensitivity
Value INT.
24
Cell 5 sensitivity
Value INT.
25
Cell 6 sensitivity
Value INT.
26
Cell 7 sensitivity
Value INT.
27
Cell 8 sensitivity
Value INT.
28
Weight division value
See the relevant table
29
Decimals
See the relevant table
30
Fixed tare (MSB)
Value INT. - Most significant word
31
Fixed tare (LSB)
Value INT. - Less significant word
32
Standby function
Value INT.
33
Keyboard block function
See the relevant table.
34
Password function
Value INT.
35
Peak function
See match on page 10
36
Datalogger function
See match on page 10
37
Datalogger Trigger
See match on page 10
38
Datalogger frequency
See match on page 10
39
Filter factor
See match on page 13
40
Output rate ADC
See match on page 13
41
Average reading number
Value INT.
42
Monotony time
Valore INT.
43
Oscillation time
Valore INT.
44
Oscillation range
Valore INT.
45
Full scale (MSB)
Valore INT. - Word più significativa