LOAD CELLS BALANCING
The procedure can be performed if at least 2 load cells are used, otherwise, the menu will not be
active.
AUTOMATIC PROCEDURE
To carry out the automatic balancing procedure, a weight must be provided to be moved over the various cells to
perform the calculation.
Once you have entered the menu and set the scale zero, you must set the value of the weight used for the procedure,
so that the automatic weighing calibration is also carried out at the end.
• At this point, the device will indicate to load the cell number 1.
• Load the sample weight and press the ENTER button
• Then proceed with the subsequent cells until the last one
If the loaded cell is not correct or if the procedure is not successful, the device will give an error; in
this case, correct the problem and try the procedure again.
BALANCING FACTOR
It is possible to manually modify the balancing factor, by accessing the appropriate menu and modifying the value
present in the selected cell.
The value can be set from 0.1000 to 9.9999 (the default value is 1.0000).