Test Measurement
NanoPlus Operator’s Manual
D-16
Mar 2014
Test Measurement
This is a preliminary test the instrument performs to determine the sign of the polarity to be
applied to the sample. The test measurement consists of ACF and Power Spectrum. After analysis,
the test results are displayed. For example:
• Positive Test Results: -12.5
• Negative Test Results: 15.8
• Polarity: (-) Negative
This decision is based on the Doppler Shift in the frequency. To interpret the results, assume the
sign of polarity that is to be determined as “x.” When “x” is multiplied with a positive value to get
a negative result (-12.5), “x” has to be negative. Similarly, when “x” is multiplied with a negative
value to get a positive result (15.8), “x” has to be negative. As a result, the sign of polarity is
negative (-).