Maintenance
NanoPlus Operator’s Manual
5-16
Mar 2014
8.
Insert electrode A into the cell, tilting the cell slightly so that air bubbles do not enter the cell,
then turn knob A clockwise to install electrode A.
Electrode A
Knob A
Do not confuse these sealing rubbers with those used for the other measuring cells.
Flat Surface Cell
This section describes how to maintain the Flat Surface Cell for Zeta Potential.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
Assembled
1
Electrode (2)
9
Luer fitting
2
Cell holder
10
Fixing nut (12)
3
O-ring (4)
11
Teflon sheet (and silicone sheet)
4
Cell block
12
Flat surface cell
5
Cell seal, translucent (2)
13
Sample fixing block
6
Sample sealing block
14
Decompression cap
7
Silicone tube
15
O-ring
8
Plug (2)
16
Clamping knob
• The Flat surface cell is coated with acrylamide; normal temperatures are recommended for
its use.