pH Maintenance Commands
NanoPlus Operator’s Manual
A-6
Mar 2014
pH Calibration
Displays the pH Calibration dialog, enabling you to calibrate the
pH electrode. Use commercially available standard solutions. A
standard solution of pH6.86 (phosphate) is necessary for the 2- and
3-point calibrations.
Calibration Points
Select 2 or 3 for the number of standards you are using.
Buffer [n]
Use these fields to select the standard solutions you are using for
the calibrations.
Results
Displays calibration results.
Registration
Saves the calibration results.
Refer to
for a step-by-
step procedure.