Maintenance
NanoPlus Operator’s Manual
5-20
Mar 2014
5.
Set the sample and teflon sheet (or silicone sheet) on the Flat Surface Cell.
Sample
Teflon sheet
Sample size
:
• Minimum: 33 x 14 mm
• Maximum: 37 x 16 x 5 mm (recommended)
Sheet selection
:
• Teflon: normal samples
• Silicone: samples of low strength
6.
Rotate the clamping knob of the sample fixing block counterclockwise until it stops; mount it
on the sample sealing block. Fix the sample fixing block with the fixing nut. Then, rotate the
clamping knob clockwise to make the sample contact closely with the Flat Surface Cell.
Loosen the decompression cap to let the air out.
Clamping knob
Decompression cap
Sample fixing block
Sample sealing block
This portion moves downward
when the clamping knob is
rotated clockwise, and vice versa.