NanoPlus Operator’s Manual
Maintenance
Mar 2014
5-17
Disassembling the Flat Surface Cell
1.
Drain the sample solution from the cell. To do this, attach the silicone tube to one of the inlets
and a syringe to the other inlet. Inject air into the cell.
Silicone tube
Luer fitting
2.
Using the special tool, remove the fixing nuts, and then remove the sample fixing block.
Special tool
Cell block
Fixing nut
Sample fixing block
3.
First, remove the fixing nuts, then the teflon (or silicone) sheet
Sample
Teflon sheet
.
4.
Remove the fixing nuts, then remove the sample sealing block and cell seals on both sides.
Cell seal
Sample sealing block
Plane cell