HPVA Series Operator’s Manual
Adjusting the Metering Valves
Mar 2011
6-13
7. On the Manual Control screen, open the gas line valve, for example, valve 6 for Gas Line 1.
8. Slightly open the metering valve.
9. Observe the Manifold pressure (
Pm
) shown on the Manual Control screen. The pressure should
rise approximately 1 bar every 10 to 20 seconds.
10. Adjust the metering valve as necessary to obtain the appropriate gas flow.
11. Close the valves on the Manual Control screen, then open valve 2 to vent.