System Configuration Window
HPVA Series Operator’s Manual
5-8
Mar 2011
Dosing Tab
Dosing Mode
Select on of three dosing mode options from the drop-down list:
Standard
The system will dose the sample to the desired pressure points
listed in the System Step Isotherm window without evacuating the
sample tube between points.
Cycle
The system will dose the manifold to the desired pressure
points listed in the System Step Isotherm window and vent the
sample after each pressure point – the Low and High Pressure
Factors will be ignored.
Selective Adsorption
The system will dose the manifold to the desired pressure
points listed in the System Step Isotherm window and will
evacuate the sample tube between points and trigger an
external device to collect said evacuated gas for composition
analysis.
Cycle Mode
Vent Time
Enter the amount of time the vent will be open after each
pressure step when using the Cycle Dosing Mode.
Содержание HPVA Series
Страница 1: ...HPVA Series High Pressure Volumetric Analyzer Operator s Manual V4 0 P02 42801 01 March 2011...
Страница 4: ......
Страница 8: ...Table of Contents HPVA Series Operator s Manual iv Mar 2011...
Страница 20: ...Specifications HPVA Series Operator s Manual 1 12 Mar 2011...
Страница 52: ...Definitions of Fields in the System Windows HPVA Series Operator s Manual 2 32 Mar 2011...
Страница 66: ...Displaying Experiment Data HPVA Series Operator s Manual 3 14 Mar 2011...
Страница 70: ...Setting the Sample Bath Temperature HPVA Series Operator s Manual 4 4 Mar 2011...
Страница 98: ...Ordering Informatioin HPVA Series Operator s Manual 7 2 Mar 2011...
Страница 120: ...Analyzing the Results HPVA Series Operator s Manual B 18 Mar 2011...
Страница 124: ...Index HPVA Series Operator s Manual Index 4 Mar 2011...